Abstract
Purpose: This paper describes a feasibility study for developing a 35-MHz high-frequency ultrasound computed-tomography (HFUCT) system for imaging rat testicles. Methods: The performances of two kinds of HFUCT-attenuation and sound-speed UCT-based on transmission and pulse-echo modes were investigated in this study. Experiments were carried out using phantoms and actual rat testicles in vitro. HFUCT images were reconstructed using a filtered backprojection algorithm. Results: The phantom experimental results indicated that all types of HFUCT can determine the dimensions of a plastic cylinder with a diameter of 500 μm. Compared to sound-speed HFUCT, attenuation HFUCT exhibited a better performance in recognizing a tiny sclerosed region in a gelatin phantom. Therefore, the in vitro testicular experiments were performed using attenuation HFUCT based on transmission and pulse-echo modes. The experimentally measured attenuation coefficient and sound speed for healthy rat testicles were 2.92 ± 0.25 dBmm and 1537 ± 25 ms, respectively. Conclusions: A homogeneous texture was evident for healthy testicles using both modes. An artificial sclerosed tumor could also be clearly observed using two- and three-dimensional attenuation HFUCT in both modes. However, an object artifact was apparent in pulse-echo mode because of ultrasound beam refraction. All of the obtained experimental results indicate the potential of using HFUCT as a novel tool for monitoring the preclinical responses of testicular tumors in small animals.
Original language | English |
---|---|
Article number | 012902 |
Journal | Medical Physics |
Volume | 41 |
Issue number | 1 |
DOIs | |
Publication status | Published - 2014 Jan |
All Science Journal Classification (ASJC) codes
- Biophysics
- Radiology Nuclear Medicine and imaging