T1 - Effect of Off-Axis Fluoroscopy Imaging on Two-Dimensional Kinematics in the Lumbar Spine
T2 - A Dynamic in Vitro Validation Study
AU - Zhao, Kristin D.
AU - Ben-Abraham, Ephraim I.
AU - Magnuson, Dixon J.
AU - Camp, Jon J.
AU - Berglund, Lawrence J.
AU - An, Kai Nan
AU - Bronfort, Gert
AU - Gay, Ralph E.
N1 - Publisher Copyright:
© 2016 by ASME.
PY - 2016/5/1
Y1 - 2016/5/1
N2 - Spine intersegmental motion parameters and the resultant regional patterns may be useful for biomechanical classification of low back pain (LBP) as well as assessing the appropriate intervention strategy. Because of its availability and reasonable cost, two-dimensional (2D) fluoroscopy has great potential as a diagnostic and evaluative tool. However, the technique of quantifying intervertebral motion in the lumbar spine must be validated, and the sensitivity assessed. The purpose of this investigation was to (1) compare synchronous fluoroscopic and optoelectronic measures of intervertebral rotations during dynamic flexion-extension movements in vitro and (2) assess the effect of C-arm rotation to simulate off-axis patient alignment on intervertebral kinematics measures. Six cadaveric lumbar-sacrum specimens were dissected, and active marker optoelectronic sensors were rigidly attached to the bodies of L2-S1. Fluoroscopic sequences and optoelectronic kinematic data (0.15-mm linear, 0.17-0.20 deg rotational, accuracy) were obtained simultaneously. After images were obtained in a true sagittal plane, the image receptor was rotated in 5 deg increments (posterior oblique angulations) from 5 deg to 15 deg. Quantitative motion analysis (qma) software was used to determine the intersegmental rotations from the fluoroscopic images. The mean absolute rotation differences between optoelectronic values and dynamic fluoroscopic values were less than 0.5 deg for all the motion segments at each off-axis fluoroscopic rotation and were not significantly different (P > 0.05) for any of the off-axis rotations of the fluoroscope. Small misalignments of the lumbar spine relative to the fluoroscope did not introduce measurement variation in relative segmental rotations greater than that observed when the spine and fluoroscope were perpendicular to each other, suggesting that fluoroscopic measures of relative segmental rotation during flexion-extension are likely robust, even when patient alignment is not perfect.
AB - Spine intersegmental motion parameters and the resultant regional patterns may be useful for biomechanical classification of low back pain (LBP) as well as assessing the appropriate intervention strategy. Because of its availability and reasonable cost, two-dimensional (2D) fluoroscopy has great potential as a diagnostic and evaluative tool. However, the technique of quantifying intervertebral motion in the lumbar spine must be validated, and the sensitivity assessed. The purpose of this investigation was to (1) compare synchronous fluoroscopic and optoelectronic measures of intervertebral rotations during dynamic flexion-extension movements in vitro and (2) assess the effect of C-arm rotation to simulate off-axis patient alignment on intervertebral kinematics measures. Six cadaveric lumbar-sacrum specimens were dissected, and active marker optoelectronic sensors were rigidly attached to the bodies of L2-S1. Fluoroscopic sequences and optoelectronic kinematic data (0.15-mm linear, 0.17-0.20 deg rotational, accuracy) were obtained simultaneously. After images were obtained in a true sagittal plane, the image receptor was rotated in 5 deg increments (posterior oblique angulations) from 5 deg to 15 deg. Quantitative motion analysis (qma) software was used to determine the intersegmental rotations from the fluoroscopic images. The mean absolute rotation differences between optoelectronic values and dynamic fluoroscopic values were less than 0.5 deg for all the motion segments at each off-axis fluoroscopic rotation and were not significantly different (P > 0.05) for any of the off-axis rotations of the fluoroscope. Small misalignments of the lumbar spine relative to the fluoroscope did not introduce measurement variation in relative segmental rotations greater than that observed when the spine and fluoroscope were perpendicular to each other, suggesting that fluoroscopic measures of relative segmental rotation during flexion-extension are likely robust, even when patient alignment is not perfect.
UR - http://www.scopus.com/inward/record.url?scp=84971291463&partnerID=8YFLogxK
UR - http://www.scopus.com/inward/citedby.url?scp=84971291463&partnerID=8YFLogxK
U2 - 10.1115/1.4032995
DO - 10.1115/1.4032995
M3 - Article
C2 - 26974192
AN - SCOPUS:84971291463
SN - 0148-0731
VL - 138
JO - Journal of Biomechanical Engineering
JF - Journal of Biomechanical Engineering
IS - 5
M1 - 054502
ER -