Abstract
Recently, endolaryngeal sonography at frequencies ranging from 10 to 30 MHz has been found to be useful in diagnosing diseases of the vocal folds (VFs). However, image resolution can be further improved by ultrasound at higher frequencies, necessitating the measurement of high-frequency acoustic properties of VF tissue. The ultrasonic parameters of integrated backscatter, sound velocity, and frequency-dependent attenuation coefficient were measured in both the lamina propria (LP) and vocalis muscle (VM) of human VFs using a 47 MHz high-frequency ultrasonic transducer. The integrated backscatter was -173.44±6.14 (mean± s.d.) and -195.13±3.58 dB in the LP and VM, respectively, the sound velocity was 1667.68±44.9 and 1595.07±39.33 ms, and the attenuation coefficient at 47 MHz was 8.28±1.72 and 7.17±1.30 dBmm. The difference between these ultrasonic parameters may be attributed to variations in the structure and fiber concentrations in VF tissue. These results could serve as a useful clinical reference for the further development of high-frequency ultrasound devices for endolarynx sonography applications.
Original language | English |
---|---|
Pages (from-to) | 1827-1832 |
Number of pages | 6 |
Journal | Journal of the Acoustical Society of America |
Volume | 122 |
Issue number | 3 |
DOIs | |
Publication status | Published - 2007 |
All Science Journal Classification (ASJC) codes
- Arts and Humanities (miscellaneous)
- Acoustics and Ultrasonics