Influence of radial head prosthetic design on radiocapitellar joint contact mechanics

  • Dipit Sahu
  • , David M. Holmes
  • , James S. Fitzsimmons
  • , Andrew R. Thoreson
  • , Lawrence J. Berglund
  • , Kai Nan An
  • , Shawn W. O'Driscoll

Research output: Contribution to journalArticlepeer-review

Abstract

Hypothesis: Our aim was to test whether anatomically designed metallic radial head implants could better reproduce native radiocapitellar contact pressure and areas than nonanatomic implants. Methods: The distal humerus and proximal radius from 6 cadaveric upper extremities were serially tested in supination with 100 N of compression force at 4 angles of flexion (0°, 30°, 60°, and 90°). By use of a thin flexible pressure transducer, contact pressures and areas were measured for the native radial head, an anatomic implant, a nonanatomic circular monopolar implant, and a bipolar nonanatomic implant. The data (mean contact pressure and mean contact area) were modeled using a 2-factor repeated-measures analysis of variance with P ≤ .05 considered to be significant. Results: The mean contact areas for the prosthetic radial heads were significantly less than those seen with the intact radial heads at every angle tested (P < .01). The mean contact pressures increased significantly with all prosthetic radial head types as compared with the native head. The mean contact pressures increased by 29% with the anatomic prosthesis, 230% with the monopolar prosthesis, and 220% with the bipolar prosthesis. Peak pressures of more than 5 MPa were more commonly observed with both the monopolar and bipolar prostheses than with the anatomic or native radial heads. Conclusions: The geometry of radial head implants strongly influences their contact characteristics. In a direct radius-to-capitellum axial loading experiment, an anatomically designed radial head prosthesis had lower and more evenly distributed contact pressures than the nonanatomic implants that were tested.

Original languageEnglish
Pages (from-to)456-462
Number of pages7
JournalJournal of Shoulder and Elbow Surgery
Volume23
Issue number4
DOIs
Publication statusPublished - 2014 Apr

All Science Journal Classification (ASJC) codes

  • Surgery
  • Orthopedics and Sports Medicine

Fingerprint

Dive into the research topics of 'Influence of radial head prosthetic design on radiocapitellar joint contact mechanics'. Together they form a unique fingerprint.

Cite this