Abstract
Hypothesis: Our aim was to test whether anatomically designed metallic radial head implants could better reproduce native radiocapitellar contact pressure and areas than nonanatomic implants. Methods: The distal humerus and proximal radius from 6 cadaveric upper extremities were serially tested in supination with 100 N of compression force at 4 angles of flexion (0°, 30°, 60°, and 90°). By use of a thin flexible pressure transducer, contact pressures and areas were measured for the native radial head, an anatomic implant, a nonanatomic circular monopolar implant, and a bipolar nonanatomic implant. The data (mean contact pressure and mean contact area) were modeled using a 2-factor repeated-measures analysis of variance with P ≤ .05 considered to be significant. Results: The mean contact areas for the prosthetic radial heads were significantly less than those seen with the intact radial heads at every angle tested (P < .01). The mean contact pressures increased significantly with all prosthetic radial head types as compared with the native head. The mean contact pressures increased by 29% with the anatomic prosthesis, 230% with the monopolar prosthesis, and 220% with the bipolar prosthesis. Peak pressures of more than 5 MPa were more commonly observed with both the monopolar and bipolar prostheses than with the anatomic or native radial heads. Conclusions: The geometry of radial head implants strongly influences their contact characteristics. In a direct radius-to-capitellum axial loading experiment, an anatomically designed radial head prosthesis had lower and more evenly distributed contact pressures than the nonanatomic implants that were tested.
| Original language | English |
|---|---|
| Pages (from-to) | 456-462 |
| Number of pages | 7 |
| Journal | Journal of Shoulder and Elbow Surgery |
| Volume | 23 |
| Issue number | 4 |
| DOIs | |
| Publication status | Published - 2014 Apr |
All Science Journal Classification (ASJC) codes
- Surgery
- Orthopedics and Sports Medicine
Fingerprint
Dive into the research topics of 'Influence of radial head prosthetic design on radiocapitellar joint contact mechanics'. Together they form a unique fingerprint.Cite this
- APA
- Author
- BIBTEX
- Harvard
- Standard
- RIS
- Vancouver