T1 - Lateral stability of proximal interphalangeal joint replacement
AU - Minamikawa, Yoshitaka
AU - Imaeda, Toshihiko
AU - Amadio, Peter C.
AU - Linscheid, Ronald L.
AU - Cooney, William P.
AU - An, Kai Nan
N1 - Funding Information:
Supported by Research Grant AR17172 from the National Institutes of Health.
PY - 1994/11
Y1 - 1994/11
N2 - The lateral stability of the cadaver proximal interphalangeal joint was studied with an electromagnetic tracking system before and after implant replacement. Ten middle fingers were retrieved and randomly divided into two groups for joint replacement with either a silicone rubber flexible finger joint implant or a custom surface replacement prosthesis. Each finger was mounted on a fixture that allowed loading of the proximal interphalangeal joint with tension through the flexor, extensor, and intrinsic tendons. For the intact proximal interphalangeal joint under lateral stress, lateral angulation averaged 4° in extension and 8° with the joint in 60° of flexion. Although lateral angulation increased after surface replacement prosthesis implantation when the joint was flexed more than 20°, comparison with the intact joint showed no statistical difference. With the silicone implant, lateral angulation became more pronounced even in the extended position and showed a significant difference when the proximal interphalangeal joint was flexed more than 20°.
AB - The lateral stability of the cadaver proximal interphalangeal joint was studied with an electromagnetic tracking system before and after implant replacement. Ten middle fingers were retrieved and randomly divided into two groups for joint replacement with either a silicone rubber flexible finger joint implant or a custom surface replacement prosthesis. Each finger was mounted on a fixture that allowed loading of the proximal interphalangeal joint with tension through the flexor, extensor, and intrinsic tendons. For the intact proximal interphalangeal joint under lateral stress, lateral angulation averaged 4° in extension and 8° with the joint in 60° of flexion. Although lateral angulation increased after surface replacement prosthesis implantation when the joint was flexed more than 20°, comparison with the intact joint showed no statistical difference. With the silicone implant, lateral angulation became more pronounced even in the extended position and showed a significant difference when the proximal interphalangeal joint was flexed more than 20°.
UR - http://www.scopus.com/inward/record.url?scp=0028116209&partnerID=8YFLogxK
UR - http://www.scopus.com/inward/citedby.url?scp=0028116209&partnerID=8YFLogxK
U2 - 10.1016/0363-5023(94)90116-3
DO - 10.1016/0363-5023(94)90116-3
M3 - Article
C2 - 7876481
AN - SCOPUS:0028116209
SN - 0363-5023
VL - 19
SP - 1050
EP - 1054
JO - Journal of Hand Surgery
JF - Journal of Hand Surgery
IS - 6
ER -