On Measurement Uncertainty of a Vortex Flowmeter

Jiun Jih J. Miau, Chen Fan Yeh, Chih Chung Hu, Jung Hua Chou

Research output: Contribution to conferencePaperpeer-review

Abstract

Measurement uncertainty of a T-shaped flowmeter was examined with a gas flow calibration facility, for Reynolds numbers, Re(h in a range of 2.56xl04-1.56xl05. The total uncertainty interval of the Strouhal values deduced from a Fourier spectral analysis was ±0.745%, and the linearity of the data was ±1.78%. Meanwhile, a scheme based on pulses counting in reference to the analog signals measured gave the total uncertainty interval of the meter factor as ±0.568% with the linearity ±1.89%.

Original languageEnglish
Pages349-354
Number of pages6
Publication statusPublished - 2004
Event12th Conference on Flow Measurement, FLOMEKO 2004 - Beijing, China
Duration: 2004 Sept 142004 Sept 17

Conference

Conference12th Conference on Flow Measurement, FLOMEKO 2004
Country/TerritoryChina
CityBeijing
Period04-09-1404-09-17

All Science Journal Classification (ASJC) codes

  • Fluid Flow and Transfer Processes

Fingerprint

Dive into the research topics of 'On Measurement Uncertainty of a Vortex Flowmeter'. Together they form a unique fingerprint.