T1 - Rapid and simple quantitative measurement of α-fetoprotein by combining immunochromatographic strip test and artificial neural network image analysis system
AU - Chuang, Louise
AU - Hwang, Jeng Yang
AU - Chang, Hen Chang
AU - Chang, Fong Ming
AU - Jong, Shiang Bin
N1 - Copyright:
Copyright 2008 Elsevier B.V., All rights reserved.
PY - 2004/10
Y1 - 2004/10
N2 - Background: Quantitative immunochromatographic strip (ICS) assay can facilitate clinical diagnosis by providing more information than traditional qualitative or semiquantitative strip assay. Methods: We constructed a human serum α-fetoprotein (AFP) measurement system by combining semiquantitative ICS tests and artificial neural network (ANN) image analysis system [immunochromatographic strip analyzed by artificial neural network image analysis system (IAIS)]. After ICS tests completed, the AFP concentration can be obtained from analysis of IAIS software. The equipment required are commercial semiquantitative AFP dipstick, optical scanner, centrifuge (option), personal computer, and software of IAIS. Results: The serum AFP concentrations measured by IAIS were strongly correlated (r=0.9971) with that by RIA. Using Bland-Altman analysis, the IAIS achieved clinical acceptable limits of agreement in comparison with RIA. The within-run precision of IAIS, expressed as coefficients of variation (CV), at 81.7 ng/ml was 1.50% and, at 244.4 ng/ml, was 1.09%. The measurement of serum AFP by IAIS can be completed in 20 min. Conclusions: The newly constructed quantitative immunochromatographic strip assay (IAIS) is a simple, rapid, and reliable method for serum AFP measurement. With the simple equipment required, the IAIS can be performed outside the laboratory and is ideal for outpatient or point-of-care AFP testing.
AB - Background: Quantitative immunochromatographic strip (ICS) assay can facilitate clinical diagnosis by providing more information than traditional qualitative or semiquantitative strip assay. Methods: We constructed a human serum α-fetoprotein (AFP) measurement system by combining semiquantitative ICS tests and artificial neural network (ANN) image analysis system [immunochromatographic strip analyzed by artificial neural network image analysis system (IAIS)]. After ICS tests completed, the AFP concentration can be obtained from analysis of IAIS software. The equipment required are commercial semiquantitative AFP dipstick, optical scanner, centrifuge (option), personal computer, and software of IAIS. Results: The serum AFP concentrations measured by IAIS were strongly correlated (r=0.9971) with that by RIA. Using Bland-Altman analysis, the IAIS achieved clinical acceptable limits of agreement in comparison with RIA. The within-run precision of IAIS, expressed as coefficients of variation (CV), at 81.7 ng/ml was 1.50% and, at 244.4 ng/ml, was 1.09%. The measurement of serum AFP by IAIS can be completed in 20 min. Conclusions: The newly constructed quantitative immunochromatographic strip assay (IAIS) is a simple, rapid, and reliable method for serum AFP measurement. With the simple equipment required, the IAIS can be performed outside the laboratory and is ideal for outpatient or point-of-care AFP testing.
UR - http://www.scopus.com/inward/record.url?scp=4444308442&partnerID=8YFLogxK
UR - http://www.scopus.com/inward/citedby.url?scp=4444308442&partnerID=8YFLogxK
U2 - 10.1016/j.cccn.2004.05.010
DO - 10.1016/j.cccn.2004.05.010
M3 - Article
C2 - 15369740
AN - SCOPUS:4444308442
SN - 0009-8981
VL - 348
SP - 87
EP - 93
JO - Clinica Chimica Acta
JF - Clinica Chimica Acta
IS - 1-2
ER -