T1 - Stem Length and Neck Resection on Fixation Strength of Press-Fit Radial Head Prosthesis
T2 - An In Vitro Model
AU - Moon, Jun Gyu
AU - Shukla, Dave R.
AU - Fitzsimmons, James S.
AU - An, Kai Nan
AU - O'Driscoll, Shawn W.
N1 - Funding Information:
This study was funded by the Mayo Foundation. The authors would like to disclose that the senior author (S.O.D.) and the Mayo Foundation receive royalties from commercial entities related to the subject of this article (Acumed, LLC; Wright Medical Group, N.V.). This study was not supported by any outside funding or grants. The Mayo Clinic Institutional Review Board that convened on December 17, 2010 approved the project, entitled “Prosthetic Radial Head Stability” (Institutional Review Board Protocol Number 10-008186). The authors also gratefully acknowledge the editorial assistance provided by Anthony M. Vaichinger during the revision of the manuscript.
Funding Information:
This study was funded by the Mayo Foundation . The authors would like to disclose that the senior author (S.O.D.) and the Mayo Foundation receive royalties from commercial entities related to the subject of this article (Acumed, LLC; Wright Medical Group, N.V.). This study was not supported by any outside funding or grants. The Mayo Clinic Institutional Review Board that convened on December 17, 2010 approved the project, entitled “Prosthetic Radial Head Stability” (Institutional Review Board Protocol Number 10-008186). The authors also gratefully acknowledge the editorial assistance provided by Anthony M. Vaichinger during the revision of the manuscript.
Publisher Copyright:
© 2019 American Society for Surgery of the Hand
PY - 2019/12
Y1 - 2019/12
N2 - Purpose: Various radial head prosthesis designs are currently in use. Few studies compare different prosthetic designs. We hypothesized that increasing a cementless implant stem's length would reduce stem–bone micromotion, with both short and long neck cuts. We also hypothesized that a minimum stem length might be required for the initial fixation strength of a press-fit implant. Methods: In 16 fresh-frozen cadaveric elbows (8 pairs), the radial head and neck were cut either 10 or 21 mm below the top of the head. Modular cementless stems were inserted and sequentially lengthened in 5-mm increments. Micromotion under eccentric loading was tested after each incremental change. Results: Incremental lengthening of the prosthetic stem and the amount of neck resection (10-mm cut vs 21-mm cut) both had a significant effect on micromotion. After a 10-mm radial head–neck resection, we observed a significant decrease in micromotion with stem lengths of 25 mm or greater, whereas with 21 mm of neck resection there was no further reduction in micromotion with increased stem length. These differences can be explained, at least in part, by the concept of the cantilever quotient: the ratio of the head–neck length outside the bone to the total length of the implant. Conclusions: The length of the stem affects the initial stability of press-fit radial head prostheses when the level of head and neck resection is at the minimum (ie, 10 mm) for currently available prosthetic designs. At this resection level, stems 25 mm or greater had significantly higher initial stability, but all stem lengths tested had mean micromotion values within the threshold for bone ingrowth. Clinical relevance: The length of a radial head prosthetic stem affects the initial stability of press-fit radial head prostheses when the level of head and neck resection is at the minimum (ie, 10 mm) for currently available prosthetic designs.
AB - Purpose: Various radial head prosthesis designs are currently in use. Few studies compare different prosthetic designs. We hypothesized that increasing a cementless implant stem's length would reduce stem–bone micromotion, with both short and long neck cuts. We also hypothesized that a minimum stem length might be required for the initial fixation strength of a press-fit implant. Methods: In 16 fresh-frozen cadaveric elbows (8 pairs), the radial head and neck were cut either 10 or 21 mm below the top of the head. Modular cementless stems were inserted and sequentially lengthened in 5-mm increments. Micromotion under eccentric loading was tested after each incremental change. Results: Incremental lengthening of the prosthetic stem and the amount of neck resection (10-mm cut vs 21-mm cut) both had a significant effect on micromotion. After a 10-mm radial head–neck resection, we observed a significant decrease in micromotion with stem lengths of 25 mm or greater, whereas with 21 mm of neck resection there was no further reduction in micromotion with increased stem length. These differences can be explained, at least in part, by the concept of the cantilever quotient: the ratio of the head–neck length outside the bone to the total length of the implant. Conclusions: The length of the stem affects the initial stability of press-fit radial head prostheses when the level of head and neck resection is at the minimum (ie, 10 mm) for currently available prosthetic designs. At this resection level, stems 25 mm or greater had significantly higher initial stability, but all stem lengths tested had mean micromotion values within the threshold for bone ingrowth. Clinical relevance: The length of a radial head prosthetic stem affects the initial stability of press-fit radial head prostheses when the level of head and neck resection is at the minimum (ie, 10 mm) for currently available prosthetic designs.
UR - http://www.scopus.com/inward/record.url?scp=85065541923&partnerID=8YFLogxK
UR - http://www.scopus.com/inward/citedby.url?scp=85065541923&partnerID=8YFLogxK
U2 - 10.1016/j.jhsa.2019.03.002
DO - 10.1016/j.jhsa.2019.03.002
M3 - Article
C2 - 31101434
AN - SCOPUS:85065541923
SN - 0363-5023
VL - 44
SP - 1098.e1-1098.e8
JO - Journal of Hand Surgery
JF - Journal of Hand Surgery
IS - 12
ER -