The relation between micromotion and screw fixation in acetabular cup

Jui Ting Hsu, Kuo-An Lai, Qingshan Chen, Mark E. Zobitz, Heng Li Huang, Kai Nan An, Chih-Han Chang

Research output: Contribution to journalArticle

35 Citations (Scopus)

Abstract

One of the major causes inducing loosening in the cementless acetabular cup implanting is its insufficient initial stability. In this study, three-dimensional finite element models of the pelvis and acetabular components were developed to investigate the relationship between relative micromotion, initial stability, and screw fixation under six daily activity loadings. A commercial available hemispheric cup with five screw holes was used as the target acetabular cup. The simulation results showed that if screws were placed closed together, when the screw number increased from 1 to 5, the peak micromotion decreased less than 14%, from 126.5 to 108.8 μm, while the stable region, micromotion less than 28 μm, enlarged only by 40%, from 46.1% to 64.7%. However, if the screw could be placed near the cup rim, a single rim screw, 202.1 μm micromotion, could provide better stability than that of four dome screws, 209.6 μm micromotion, placed closed together. To conclude, multiple cup screws should be placed near cup rim and as separate as possible to enlarge the stable region and reduce the peak micromotion between cup and acetabulum.

Original languageEnglish
Pages (from-to)34-41
Number of pages8
JournalComputer Methods and Programs in Biomedicine
Volume84
Issue number1
DOIs
Publication statusPublished - 2006 Oct 1

Fingerprint

Acetabulum
Pelvis
Domes

All Science Journal Classification (ASJC) codes

  • Software
  • Computer Science Applications
  • Health Informatics

Cite this

Hsu, Jui Ting ; Lai, Kuo-An ; Chen, Qingshan ; Zobitz, Mark E. ; Huang, Heng Li ; An, Kai Nan ; Chang, Chih-Han. / The relation between micromotion and screw fixation in acetabular cup. In: Computer Methods and Programs in Biomedicine. 2006 ; Vol. 84, No. 1. pp. 34-41.
@article{17c9728f38294e81b546922b7050ddf0,
title = "The relation between micromotion and screw fixation in acetabular cup",
abstract = "One of the major causes inducing loosening in the cementless acetabular cup implanting is its insufficient initial stability. In this study, three-dimensional finite element models of the pelvis and acetabular components were developed to investigate the relationship between relative micromotion, initial stability, and screw fixation under six daily activity loadings. A commercial available hemispheric cup with five screw holes was used as the target acetabular cup. The simulation results showed that if screws were placed closed together, when the screw number increased from 1 to 5, the peak micromotion decreased less than 14{\%}, from 126.5 to 108.8 μm, while the stable region, micromotion less than 28 μm, enlarged only by 40{\%}, from 46.1{\%} to 64.7{\%}. However, if the screw could be placed near the cup rim, a single rim screw, 202.1 μm micromotion, could provide better stability than that of four dome screws, 209.6 μm micromotion, placed closed together. To conclude, multiple cup screws should be placed near cup rim and as separate as possible to enlarge the stable region and reduce the peak micromotion between cup and acetabulum.",
author = "Hsu, {Jui Ting} and Kuo-An Lai and Qingshan Chen and Zobitz, {Mark E.} and Huang, {Heng Li} and An, {Kai Nan} and Chih-Han Chang",
year = "2006",
month = "10",
day = "1",
doi = "10.1016/j.cmpb.2006.08.002",
language = "English",
volume = "84",
pages = "34--41",
journal = "Computer Methods and Programs in Biomedicine",
issn = "0169-2607",
publisher = "Elsevier Ireland Ltd",
number = "1",

}

The relation between micromotion and screw fixation in acetabular cup. / Hsu, Jui Ting; Lai, Kuo-An; Chen, Qingshan; Zobitz, Mark E.; Huang, Heng Li; An, Kai Nan; Chang, Chih-Han.

In: Computer Methods and Programs in Biomedicine, Vol. 84, No. 1, 01.10.2006, p. 34-41.

Research output: Contribution to journalArticle

TY - JOUR

T1 - The relation between micromotion and screw fixation in acetabular cup

AU - Hsu, Jui Ting

AU - Lai, Kuo-An

AU - Chen, Qingshan

AU - Zobitz, Mark E.

AU - Huang, Heng Li

AU - An, Kai Nan

AU - Chang, Chih-Han

PY - 2006/10/1

Y1 - 2006/10/1

N2 - One of the major causes inducing loosening in the cementless acetabular cup implanting is its insufficient initial stability. In this study, three-dimensional finite element models of the pelvis and acetabular components were developed to investigate the relationship between relative micromotion, initial stability, and screw fixation under six daily activity loadings. A commercial available hemispheric cup with five screw holes was used as the target acetabular cup. The simulation results showed that if screws were placed closed together, when the screw number increased from 1 to 5, the peak micromotion decreased less than 14%, from 126.5 to 108.8 μm, while the stable region, micromotion less than 28 μm, enlarged only by 40%, from 46.1% to 64.7%. However, if the screw could be placed near the cup rim, a single rim screw, 202.1 μm micromotion, could provide better stability than that of four dome screws, 209.6 μm micromotion, placed closed together. To conclude, multiple cup screws should be placed near cup rim and as separate as possible to enlarge the stable region and reduce the peak micromotion between cup and acetabulum.

AB - One of the major causes inducing loosening in the cementless acetabular cup implanting is its insufficient initial stability. In this study, three-dimensional finite element models of the pelvis and acetabular components were developed to investigate the relationship between relative micromotion, initial stability, and screw fixation under six daily activity loadings. A commercial available hemispheric cup with five screw holes was used as the target acetabular cup. The simulation results showed that if screws were placed closed together, when the screw number increased from 1 to 5, the peak micromotion decreased less than 14%, from 126.5 to 108.8 μm, while the stable region, micromotion less than 28 μm, enlarged only by 40%, from 46.1% to 64.7%. However, if the screw could be placed near the cup rim, a single rim screw, 202.1 μm micromotion, could provide better stability than that of four dome screws, 209.6 μm micromotion, placed closed together. To conclude, multiple cup screws should be placed near cup rim and as separate as possible to enlarge the stable region and reduce the peak micromotion between cup and acetabulum.

UR - http://www.scopus.com/inward/record.url?scp=33748807200&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=33748807200&partnerID=8YFLogxK

U2 - 10.1016/j.cmpb.2006.08.002

DO - 10.1016/j.cmpb.2006.08.002

M3 - Article

C2 - 16971018

AN - SCOPUS:33748807200

VL - 84

SP - 34

EP - 41

JO - Computer Methods and Programs in Biomedicine

JF - Computer Methods and Programs in Biomedicine

SN - 0169-2607

IS - 1

ER -