Wireless biopotential acquisition system for portable healthcare monitoring

W. S. Wang, H. Y. Huang, Z. C. Wu, S. C. Chen, W. F. Wang, C. F. Wu, C. H. Luo

Research output: Contribution to journalArticlepeer-review

11 Citations (Scopus)

Abstract

A complete biopotential acquisition system with an analogue front-end (AFE) chip is proposed for portable healthcare monitoring. A graphical user interface (GUI) is also implemented to display the extracted biopotential signals in real-time on a computer for patients or in a hospital via the internet for doctors. The AFE circuit defines the quality of the acquired biosignals. Thus, an AFE chip with low power consumption and a high common-mode rejection ratio (CMRR) was implemented in the TSMC 0.18-μm CMOS process. The measurement results show that the proposed AFE, with a core area of 0.1 mm2, has a CMRR of 90 dB, and power consumption of 21.6μW. Biopotential signals of electroencephalogram (EEG), electrocardiogram (ECG) and electromyogram (EMG) were measured to verify the proposed system. The board size of the proposed system is 6 cm × - 2.5 cm and the weight is 30 g. The total power consumption of the proposed system is 66 mW.

Original languageEnglish
Pages (from-to)254-261
Number of pages8
JournalJournal of Medical Engineering and Technology
Volume35
Issue number5
DOIs
Publication statusPublished - 2011 Jul

All Science Journal Classification (ASJC) codes

  • Biomedical Engineering

Fingerprint Dive into the research topics of 'Wireless biopotential acquisition system for portable healthcare monitoring'. Together they form a unique fingerprint.