摘要
The conductance catheter system is a tool to determine left ventricular (LV) volume in vivo. However, the accuracy of this method is limited by three assumptions of the equation to convert conductance to volume: the assumed homogeneity of electric field within LV, myocardium having resistive and not capacitive properties, and assumed constant parallel myocardial conductance. This study investigates weaknesses of these three assumptions. It proposes an equivalent circuit model for mouse LV as well as a calculation method for conductance and capacitance of LV. Furthermore, this study also demonstrates the importance of phase measurement in vivo experiments.
| 原文 | English |
|---|---|
| 頁(從 - 到) | 62-65 |
| 頁數 | 4 |
| 期刊 | Annual International Conference of the IEEE Engineering in Medicine and Biology - Proceedings |
| 卷 | 1 |
| 出版狀態 | Published - 2003 |
| 事件 | A New Beginning for Human Health: Proceedings of the 25th Annual International Conference of the IEEE Engineering in Medicine and Biology Society - Cancun, Mexico 持續時間: 2003 9月 17 → 2003 9月 21 |
All Science Journal Classification (ASJC) codes
- 訊號處理
- 生物醫學工程
- 電腦視覺和模式識別
- 健康資訊學
指紋
深入研究「3D Finite Element Complex Domain Numerical Models of Electric Fields in Blood and Myocardium」主題。共同形成了獨特的指紋。引用此
- APA
- Author
- BIBTEX
- Harvard
- Standard
- RIS
- Vancouver