A stabilization device that promotes the efficiency of cardiopulmonary resuscitation during ambulance transportation to the level as under non-moving conditions

Ning Ping Foo, Jer Hao Chang, Shih Bin Su, Kow Tong Chen, Ching Fa Cheng, Pei Chung Chen, Tsung Yi Lin, How Ran Guo

研究成果: Article同行評審

3 引文 斯高帕斯(Scopus)

摘要

Results: The overall effective compressions in 10 minutes were 86.4±17.5% for NM, 60.9±14.6% for MND, 69.7±22.4% for MD, and 86.6%±13.2% for MVSD (p<0.001). Whereas changes in back pain severity and physiology parameters were similar under all conditions, MVSD had the lowest no-flow fraction. Differences in effective compressions and the no-flow fraction between MVSD and NM did not reach statistical significance.

Conclusions: The use of the modified device can improve quality of CPR in a moving ambulance to a level similar to that in a non-moving condition without increasing the severity of back pain.

Background: The survival rate of patients with out-of-hospital cardiac arrest is low, and measures to improve the quality of cardiopulmonary resuscitation (CPR) during ambulance transportation are desirable. We designed a stabilization device, and in a randomized crossover trial we found performing CPR in a moving ambulance with the device (MD) could achieve better efficiency than that without the device (MND), but the efficiency was lower than that in a non-moving ambulance (NM).

Purpose: To evaluate whether a modified version of the stabilization device, can promote further the quality of CPR during ambulance transportation.

Methods: Participants of the previous study were recruited, and they performed CPR for 10 minutes in a moving ambulance with the modified version of the stabilization device (MVSD). The primary outcomes were effective chest compressions and no-flow fraction recorded by a skill-reporter manikin. The secondary outcomes included back pain, physiological parameters, and the participants' rating about the device after performing CPR.

原文English
文章編號e107960
期刊PloS one
9
發行號10
DOIs
出版狀態Published - 2014 10月 15

All Science Journal Classification (ASJC) codes

  • 多學科

指紋

深入研究「A stabilization device that promotes the efficiency of cardiopulmonary resuscitation during ambulance transportation to the level as under non-moving conditions」主題。共同形成了獨特的指紋。

引用此