TY - JOUR
T1 - Analysis of longitudinal arch supports in stabilizing the arch of the foot
AU - Kitaoka, Harold B.
AU - Luo, Zong Ping
AU - An, Kai Nan
PY - 1997/8
Y1 - 1997/8
N2 - Two orthotic devices commonly prescribed as arch supports were studied to evaluate their efficacy in stabilizing the foot. Fourteen cadaveric feet were mounted in a loading frame, and an axial load of 222, 445, or 667 N was applied while three-dimensional positions of the talus, calcaneus, navicular, and first metatarsal were monitored with a magnetic tracking system. Feet were tested with and without the use of two commonly prescribed arch supports. The two indices used to assess arch stabilization were arch height and joint rotation. Joint rotations consistently increased on load application. Significant differences were observed with Inserts 1 and 2 in metatarsal talar abduction, dorsiflexion and eversion, calcaneal talar eversion, and talar tibial dorsiflexion. Arch height significantly increased with both inserts. Contrary to previously published results, the arch supports that were studied provided measurable improvement in arch stability in a simulated standing at ease position.
AB - Two orthotic devices commonly prescribed as arch supports were studied to evaluate their efficacy in stabilizing the foot. Fourteen cadaveric feet were mounted in a loading frame, and an axial load of 222, 445, or 667 N was applied while three-dimensional positions of the talus, calcaneus, navicular, and first metatarsal were monitored with a magnetic tracking system. Feet were tested with and without the use of two commonly prescribed arch supports. The two indices used to assess arch stabilization were arch height and joint rotation. Joint rotations consistently increased on load application. Significant differences were observed with Inserts 1 and 2 in metatarsal talar abduction, dorsiflexion and eversion, calcaneal talar eversion, and talar tibial dorsiflexion. Arch height significantly increased with both inserts. Contrary to previously published results, the arch supports that were studied provided measurable improvement in arch stability in a simulated standing at ease position.
UR - http://www.scopus.com/inward/record.url?scp=0343051818&partnerID=8YFLogxK
UR - http://www.scopus.com/inward/citedby.url?scp=0343051818&partnerID=8YFLogxK
U2 - 10.1097/00003086-199708000-00036
DO - 10.1097/00003086-199708000-00036
M3 - Article
C2 - 9269181
AN - SCOPUS:0343051818
SN - 0009-921X
VL - 341
SP - 250
EP - 256
JO - Clinical Orthopaedics and Related Research
JF - Clinical Orthopaedics and Related Research
ER -