TY - JOUR
T1 - Apparatus for Comparison of Pullout Forces for Various Thoracic Stent Grafts at Varying Neck Angulations and Oversizes
AU - Chiang, Cheng Hsien
AU - Yeh, Ming Long
AU - Chen, Wei Ling
AU - Kan, Chung Dann
N1 - Publisher Copyright:
© 2016 Elsevier Inc.
PY - 2016/2/1
Y1 - 2016/2/1
N2 - Background The purpose of this study is to provide an apparatus for comparison of pullout forces for various thoracic stent grafts at varying neck angulations and oversizes. Methods An in vitro platform capable of performing pullout tests was used on stent grafts in angulated silicone tubes designed for this study (0°, 45°, 90°, and 135° with a 32-mm inner diameter) in a temperature-controlled chamber (37 ± 2°C). Three commercial stent grafts with sizes commonly used in Taiwan (Valiant: 34, 36, 38, and 40 mm; Zenith TX2: 34, 36, 38, and 40 mm; and TAG: 34, 37, and 40 mm) were used, and each size was tested 8 times for each angulation condition. Results The mean dislodgement forces (DFs) at 0° angulation within 10–20% oversize were approximately 22.7, 9.6, and 9.0 N for the Valiant, Zenith TX2, and TAG devices, respectively, whereas the mean DFs decreased by 46%, 38%, and 50% to 12.3, 5.9, and 4.5 N when the angulation reached 135°. Regression analysis shows that neck angulation was a significant factor for the Valiant and Zenith TX2 devices (P < 0.0001 and P < 0.0001, respectively) but not for the TAG device (P = 0.483). In addition, oversize and interactions between variables (angulation × oversize) exhibited significant effects on the DFs for all devices (P < 0.0001). Conclusions We successfully built up an apparatus for comparison of pullout forces for various thoracic stent grafts at varying neck angulations and oversizes. With the empirical comparative data of different brand stent grafts under various conditions shown and compared, our findings suggest that aortic neck angulation has a negative correlation with stent-graft fixation. To have better stent-graft fixation and seal in the aortic arch for thoracic endovascular aortic repair, a longer landing zone with cautiously selected oversize is a more suitable selection.
AB - Background The purpose of this study is to provide an apparatus for comparison of pullout forces for various thoracic stent grafts at varying neck angulations and oversizes. Methods An in vitro platform capable of performing pullout tests was used on stent grafts in angulated silicone tubes designed for this study (0°, 45°, 90°, and 135° with a 32-mm inner diameter) in a temperature-controlled chamber (37 ± 2°C). Three commercial stent grafts with sizes commonly used in Taiwan (Valiant: 34, 36, 38, and 40 mm; Zenith TX2: 34, 36, 38, and 40 mm; and TAG: 34, 37, and 40 mm) were used, and each size was tested 8 times for each angulation condition. Results The mean dislodgement forces (DFs) at 0° angulation within 10–20% oversize were approximately 22.7, 9.6, and 9.0 N for the Valiant, Zenith TX2, and TAG devices, respectively, whereas the mean DFs decreased by 46%, 38%, and 50% to 12.3, 5.9, and 4.5 N when the angulation reached 135°. Regression analysis shows that neck angulation was a significant factor for the Valiant and Zenith TX2 devices (P < 0.0001 and P < 0.0001, respectively) but not for the TAG device (P = 0.483). In addition, oversize and interactions between variables (angulation × oversize) exhibited significant effects on the DFs for all devices (P < 0.0001). Conclusions We successfully built up an apparatus for comparison of pullout forces for various thoracic stent grafts at varying neck angulations and oversizes. With the empirical comparative data of different brand stent grafts under various conditions shown and compared, our findings suggest that aortic neck angulation has a negative correlation with stent-graft fixation. To have better stent-graft fixation and seal in the aortic arch for thoracic endovascular aortic repair, a longer landing zone with cautiously selected oversize is a more suitable selection.
UR - http://www.scopus.com/inward/record.url?scp=84950297117&partnerID=8YFLogxK
UR - http://www.scopus.com/inward/citedby.url?scp=84950297117&partnerID=8YFLogxK
U2 - 10.1016/j.avsg.2015.10.007
DO - 10.1016/j.avsg.2015.10.007
M3 - Article
C2 - 26597245
AN - SCOPUS:84950297117
SN - 0890-5096
VL - 31
SP - 196
EP - 204
JO - Annals of Vascular Surgery
JF - Annals of Vascular Surgery
ER -