摘要
Peripheral arterial disease (PAD) is highly encountered in hemodialysis patients. The rate of cardiovascular disease-Associated hospitalization and mortality among those PAD patients is higher than non- PAD subjects. Early detection of PAD could prevent peripheral limbs problems such as numbness, limp, and amputations, especially in hemodialysis patients' foot. The ankle brachial index (ABI) is a widely used tool for PAD screening; however it has many clinical limitations and is not suitable for the patients with vascular sclerosis. Therefore, the aim of the present study is to develop a bilateral photoplethysmography (PPG) system with Van der Pol (VDP) Oscillator detector based on Burg autoregressive (AR) method for PAD estimation. The proposed system could track the bilateral symmetry or asymmetry of PPG signals and classify the grade of PAD in hemodialysis patients. The test results from 32 patients showed that the proposed system has high accuracy of classification (over 90%) for hemodialysis patients.
原文 | English |
---|---|
頁面 | 154-157 |
頁數 | 4 |
DOIs | |
出版狀態 | Published - 2014 |
事件 | IASTED International Conference on Biomedical Engineering, BioMed 2014 - Zurich, Switzerland 持續時間: 2014 6月 23 → 2014 6月 25 |
Other
Other | IASTED International Conference on Biomedical Engineering, BioMed 2014 |
---|---|
國家/地區 | Switzerland |
城市 | Zurich |
期間 | 14-06-23 → 14-06-25 |
All Science Journal Classification (ASJC) codes
- 建模與模擬