Complications of port-catheter systems in intra-arterial infusion chemotherapy

M. C. Sheen, Y. W. Wang

研究成果: Article同行評審

3 引文 斯高帕斯(Scopus)

摘要

The complications associated with the placement and use of 252 modified Port-A-Cath systems and 24 Jet Port systems (PFM, FRG) in 266 cancer patients, who received arterial infusion chemotherapy during a 31-month period, were analyzed. The Port-A-Cath (Pharmacia, USA) was connected to a Teflon catheter (Sharp, Japan) for infusion therapy of various regional malignancies. Complications necessitating reoperation are grouped in two categories: port-catheter occlusions and local wound complications. Those related to port-catheter occlusions (n=76, 28%) leading to catheter rupture occured in 55, disconnection between the port and the catheter in 16, and luxation of the port in 5. Local wound complications (n=20,7%) encountered were wound dehiscence in 12, thinning of the skin over the port in 4, puncture-site skin necrosis in 2, pocket skin necrosis in 1 and port rotation in 1. All these complications can be easily dealt with without any serious consequences. Most of the above complications can be avoided by using an adequately designed port-catheter system, proper handling by the expert, and use of a suitable heparin.

原文English
頁(從 - 到)92-97
頁數6
期刊Regional Cancer Treatment
4
發行號2
出版狀態Published - 1991 一月 1

All Science Journal Classification (ASJC) codes

  • 腫瘤科
  • 藥理

指紋

深入研究「Complications of port-catheter systems in intra-arterial infusion chemotherapy」主題。共同形成了獨特的指紋。

引用此