TY - JOUR
T1 - Computational comparison of bone cement and poly aryl-ether-ether-ketone spacer in single-segment posterior lumbar interbody fusion
T2 - a pilot study
AU - Chang, Chih Wei
AU - Chung, Yu Hsuan
AU - Chang, Chia Jung
AU - Chen, Yen Nien
AU - Li, Chun Ting
AU - Chang, Chih Han
AU - Peng, Yao Te
N1 - Funding Information:
The authors disclosed receipt of the following financial support for the research, authorship, and/or publication of this article: This work was supported by National Cheng Kung University—Show Chwan Health Care System R&D Project (NCKUSCMH10612).
Publisher Copyright:
© 2019, Australasian College of Physical Scientists and Engineers in Medicine.
PY - 2019
Y1 - 2019
N2 - Posterior lumbar interbody fusion (PLIF) with a spacer and posterior instrument (PI) via minimally invasive surgery (MIS) restores intervertebral height in degenerated disks. To align with MIS, the spacer has to be shaped with a slim geometry. However, the thin spacer increases the subsidence and migration after PLIF. This study aimed to propose a new lumbar fusion approach using bone cement to achieve a larger supporting area than that achieved by the currently used poly aryl-ether-ether-ketone (PEEK) spacer and assess the feasibility of this approach using a sawbone model. Furthermore, the mechanical responses, including the range of motion (ROM) and bone stress with the bone cement spacer were compared to those noted with the PEEK spacer by finite element (FE) simulation. An FE lumbar L3–L4 model with PEEK and bone cement spacers and PI was developed. Four fixing conditions were considered: intact lumbar L3–L4 segment, lumbar L3–L4 segment with PI, PEEK spacer plus PI, and bone cement spacer plus PI. Four kinds of 10-NM moments (flexion, extension, lateral bending, and rotation) and two different bone qualities (normal and osteoporotic) were considered. The bone cement spacer yielded smaller ROMs in extension and rotation than the PEEK spacer, while the ROMs of the bone cement spacer in flexion and lateral bending were slightly greater than with the PEEK spacer. Compared with the PEEK spacer, peak contact pressure on the superior surface of L4 with the bone cement spacer in rotation decreased by 74% (from 8.68 to 2.25 MPa) and 69.1% (from 9.1 to 2.82 MPa), respectively, in the normal and osteoporotic bone. Use of bone cement as a spacer with PI is a potential approach to decrease the bone stress in lumbar fusion and warrants further research.
AB - Posterior lumbar interbody fusion (PLIF) with a spacer and posterior instrument (PI) via minimally invasive surgery (MIS) restores intervertebral height in degenerated disks. To align with MIS, the spacer has to be shaped with a slim geometry. However, the thin spacer increases the subsidence and migration after PLIF. This study aimed to propose a new lumbar fusion approach using bone cement to achieve a larger supporting area than that achieved by the currently used poly aryl-ether-ether-ketone (PEEK) spacer and assess the feasibility of this approach using a sawbone model. Furthermore, the mechanical responses, including the range of motion (ROM) and bone stress with the bone cement spacer were compared to those noted with the PEEK spacer by finite element (FE) simulation. An FE lumbar L3–L4 model with PEEK and bone cement spacers and PI was developed. Four fixing conditions were considered: intact lumbar L3–L4 segment, lumbar L3–L4 segment with PI, PEEK spacer plus PI, and bone cement spacer plus PI. Four kinds of 10-NM moments (flexion, extension, lateral bending, and rotation) and two different bone qualities (normal and osteoporotic) were considered. The bone cement spacer yielded smaller ROMs in extension and rotation than the PEEK spacer, while the ROMs of the bone cement spacer in flexion and lateral bending were slightly greater than with the PEEK spacer. Compared with the PEEK spacer, peak contact pressure on the superior surface of L4 with the bone cement spacer in rotation decreased by 74% (from 8.68 to 2.25 MPa) and 69.1% (from 9.1 to 2.82 MPa), respectively, in the normal and osteoporotic bone. Use of bone cement as a spacer with PI is a potential approach to decrease the bone stress in lumbar fusion and warrants further research.
UR - http://www.scopus.com/inward/record.url?scp=85076301038&partnerID=8YFLogxK
UR - http://www.scopus.com/inward/citedby.url?scp=85076301038&partnerID=8YFLogxK
U2 - 10.1007/s13246-019-00832-8
DO - 10.1007/s13246-019-00832-8
M3 - Article
AN - SCOPUS:85076301038
SN - 0158-9938
JO - Australasian Physical and Engineering Sciences in Medicine
JF - Australasian Physical and Engineering Sciences in Medicine
ER -