Detection of C-reactive protein on an integrated microfluidic system by utilizing field-effect transistors and aptamers

Wei Chieh Kao, Yen Wen Chen, Chia Ho Chu, Wen Hsin Chang, Shu Chu Shiesh, Yu Lin Wang, Gwo Bin Lee

研究成果: Article同行評審

16 引文 斯高帕斯(Scopus)

摘要

Cardiovascular diseases (CVDs) cause more than 17 × 106 deaths worldwide on a yearly basis. Early diagnosis of CVDs is therefore of great need. The C-reactive protein (CRP) is an important biomarker for analyzing the risks of CVDs. In this work, CRP-specific aptamers with high sensitivity and specificity and fieldeffect- transistor (FET) devices were used to recognize and detect CRP by using an integrated microfluidic system automatically while consuming less volumes of reagents and samples (about 5 μm). In order to package the FET device into the microfluidic chip, a new method to prevent liquid leakage was proposed. Sensitive detection of CRP has been demonstrated on the developed microfluidic system. It is the first time that aptamer-FET assays could be realized on an integrated microfluidic system. Experimental results showed that the aptamer-FET assay was capable of detecting CRP with concentrations ranging from 0.625 mg/l to 10.000 mg/l, which may be promising for early diagnosis of CVDs.

原文English
文章編號044105
期刊Biomicrofluidics
11
發行號4
DOIs
出版狀態Published - 2017 7月 1

All Science Journal Classification (ASJC) codes

  • 生物醫學工程
  • 一般材料科學
  • 凝聚態物理學
  • 流體流動和轉移過程
  • 膠體和表面化學

指紋

深入研究「Detection of C-reactive protein on an integrated microfluidic system by utilizing field-effect transistors and aptamers」主題。共同形成了獨特的指紋。

引用此