Development, integration and use of an ultrahigh-strength gradient system on a human-size 3 T magnet for small animal MRI

  • Kuan Hung Cho
  • , Sheng Min Huang
  • , Chang Hoon Choi
  • , Ming Jye Chen
  • , Hsuan Han Chiang
  • , Richard P. Buschbeck
  • , Ezequiel Farrher
  • , N. Jon Shah
  • , Ruslan Garipov
  • , Ching Ping Chang
  • , Hsu Chang
  • , Li Wei Kuo

研究成果: Article同行評審

13   !!Link opens in a new tab 引文 斯高帕斯(Scopus)

摘要

This study aims to integrate an ultra-high-strength gradient coil system on a clinical 3 T magnet and demonstrate its preclinical imaging capabilities. Dedicated phantoms were used to qualitatively and quantitatively assess the performance of the gradient system. Advanced MR imaging sequences, including diffusion tensor imaging (DTI) and quantitative susceptibility mapping (QSM), were implemented and executed on an ex vivo specimen as well as in vivo rats. The DTI and QSM results on the phantom agreed well with those in the literature. Furthermore, studies on ex vivo specimens have demonstrated the applicability of DTI and QSM on our system to probe microstructural changes in a mild traumatic brain injury rat model. The feasibility of in vivo rat DTI was also demonstrated. We showed that the inserted ultra-high-strength gradient coil was successfully integrated on a clinically used magnet. After careful tuning and calibration, we verified the accuracy and quantitative preclinical imaging capability of the integrated system in phantom and in vivo rat brain experiments. This study can be essential to establish dedicated animal MRI platform on clinical MRI scanners and facilitate translational studies at clinical settings.

原文English
文章編號e0217916
期刊PloS one
14
發行號6
DOIs
出版狀態Published - 2019 6月

All Science Journal Classification (ASJC) codes

  • 多學科

指紋

深入研究「Development, integration and use of an ultrahigh-strength gradient system on a human-size 3 T magnet for small animal MRI」主題。共同形成了獨特的指紋。

引用此