Development of flow through dielectrophoresis microfluidic chips for biofuel production: Sorting and detection of microalgae with different lipid contents

Yu Luen Deng, Mei Yi Kuo, Yi Je Juang

研究成果: Article同行評審

29 引文 斯高帕斯(Scopus)

摘要

In this study, a continuous flow dielectrophoresis (DEP) microfluidic chip was fabricated and utilized to sort out the microalgae (C. vulgaris) with different lipid contents. The proposed separation scheme is to allow that the microalgae with different lipid contents experience different negative or no DEP force at the separation electrode pair under the pressure-driven flow. The microalgae that experience stronger negative DEP will be directed to the side channel while those experience less negative or no DEP force will pass through the separation electrode pair to remain in the main channel. It was found that the higher the lipid content inside the microalgae, the higher the crossover frequency. Separation of the microalgae with 13% and 21% lipid contents, and 24% and 30%-35% lipid contents was achieved at the operating frequency 7?MHz, and 10 MHz, respectively. Moreover, separation can be further verified by measurement of the fluorescence intensity of the neutral lipid inside the sorted algal cells.

原文English
文章編號064120
期刊Biomicrofluidics
8
發行號6
DOIs
出版狀態Published - 2014 12月 9

All Science Journal Classification (ASJC) codes

  • 生物醫學工程
  • 材料科學(全部)
  • 凝聚態物理學
  • 流體流動和轉移過程
  • 膠體和表面化學

指紋

深入研究「Development of flow through dielectrophoresis microfluidic chips for biofuel production: Sorting and detection of microalgae with different lipid contents」主題。共同形成了獨特的指紋。

引用此