T1 - Evaluation of Lens Hardness in Cataract Surgery using High-Frequency Ultrasonic Parameters in Vitro
AU - Huang, Chih Chung
AU - Ameri, Hossein
AU - DeBoer, Charles
AU - Rowley, Adrian P.
AU - Xu, Xiaochen
AU - Sun, Lei
AU - Wang, Shyh Hau
AU - Humayun, Mark S.
AU - Shung, K. Kirk
N1 - Funding Information:
This work was supported by the Doheny Retina Institute, Doheny Eye Institute, Department of Ophthalmology, Keck School of Medicine, University of Southern California, USA.
PY - 2007/10
Y1 - 2007/10
N2 - Ultrasonic parameters of sound velocity and frequency-dependent attenuation ranging from 25 to 45 MHz were measured for the purpose of evaluating the hardness of lenses in cataract surgery (phacoemulsification). Measurements were performed with a 35-MHz ultrasonic transducer on porcine lenses in which artificially cataracts were induced. The hardness of the cataractous lens was also evaluated by mechanical measurement of its elastic properties. The results indicated that the ultrasonic attenuation coefficients in normal porcine lenses were approximately 4.49 ± 0.05 (mean ± SD) and 6.32 ± 0.04 dB/mm at 30 and 40 MHz, respectively. The development progression of the cataracts resulted in the attenuation coefficient increasing linearly to 7.36 ± 0.25 and 11.1 ± 0.92 dB/mm, respectively, corresponding to an increase of Young's modulus from 2.6 to 101.2 kPa. The sound velocity concomitantly increased from 1639.8 ± 4.2 to 1735.6 ± 10.4 m/s. Evaluation of the relationship between the phacoemulsification energy level and ultrasonic parameters in vitro by surgeons revealed that both the attenuation coefficient and sound velocity were linearly correlated with the phacoemulsification energy (r = 0.941 and 0.915, respectively). These results showed that measuring high-frequency ultrasonic parameters provides surgeons with good capability and reproducibility for selecting the optimal energy level for phacoemulsification. (E-mail: [email protected]).
AB - Ultrasonic parameters of sound velocity and frequency-dependent attenuation ranging from 25 to 45 MHz were measured for the purpose of evaluating the hardness of lenses in cataract surgery (phacoemulsification). Measurements were performed with a 35-MHz ultrasonic transducer on porcine lenses in which artificially cataracts were induced. The hardness of the cataractous lens was also evaluated by mechanical measurement of its elastic properties. The results indicated that the ultrasonic attenuation coefficients in normal porcine lenses were approximately 4.49 ± 0.05 (mean ± SD) and 6.32 ± 0.04 dB/mm at 30 and 40 MHz, respectively. The development progression of the cataracts resulted in the attenuation coefficient increasing linearly to 7.36 ± 0.25 and 11.1 ± 0.92 dB/mm, respectively, corresponding to an increase of Young's modulus from 2.6 to 101.2 kPa. The sound velocity concomitantly increased from 1639.8 ± 4.2 to 1735.6 ± 10.4 m/s. Evaluation of the relationship between the phacoemulsification energy level and ultrasonic parameters in vitro by surgeons revealed that both the attenuation coefficient and sound velocity were linearly correlated with the phacoemulsification energy (r = 0.941 and 0.915, respectively). These results showed that measuring high-frequency ultrasonic parameters provides surgeons with good capability and reproducibility for selecting the optimal energy level for phacoemulsification. (E-mail: [email protected]).
UR - http://www.scopus.com/inward/record.url?scp=34648813642&partnerID=8YFLogxK
UR - http://www.scopus.com/inward/citedby.url?scp=34648813642&partnerID=8YFLogxK
U2 - 10.1016/j.ultrasmedbio.2007.05.002
DO - 10.1016/j.ultrasmedbio.2007.05.002
M3 - Article
C2 - 17618041
AN - SCOPUS:34648813642
SN - 0301-5629
VL - 33
SP - 1609
EP - 1616
JO - Ultrasound in Medicine and Biology
JF - Ultrasound in Medicine and Biology
IS - 10
ER -