We investigates the effect of the anchor area on the mechanical strengths of infrazygomatic mini-implants. Thirty mini-implants were divided into three types based on the material and shape: Type A (titanium alloy, 2.0 × 12 mm), Type B (stainless steel, 2.0 × 12 mm), and Type C (titanium alloy, 2.0 × 11 mm).The mini-implants were inserted at 90° and 45° into the artificial bone to a depth of 7 mm, without predrilling. The mechanical strengths [insertion torque (IT), resonance frequency (RF), and removal torque (RT)] and the anchor area were measured. We hypothesized that no correlation exists among the mechanical forces of each brand. In the 90° tests, the IT, RF, and RT of Type C (8.5 N cm, 10.2 kHz, and 6.1 N cm, respectively) were significantly higher than those of Type A (5.0 N cm, 7.7 kHz, and 4.7 N cm, respectively). In the 45° test, the RFs of Type C (9.2 kHz) was significantly higher than those of Type A (7.0 kHz) and Type B (6.7 kHz). The anchor area of the mini-implants was in the order of Type C (706 mm2) > Type B (648 mm2) > Type A (621 mm2). Type C exhibited no significant correlation in intragroup comparisons, and the hypothesis was accepted. In the 90° and 45° tests, Type C exhibited the largest anchor area and the highest mechanical strengths (IT, RF, and RT) among the three types of mini-implants. The anchor area plays a crucial role in the mechanical strength of mini-implants.
| 原文 | English |
|---|---|
| 頁(從 - 到) | 96-101 |
| 頁數 | 6 |
| 期刊 | Kaohsiung Journal of Medical Sciences |
| 卷 | 33 |
| 發行號 | 2 |
| DOIs | |
| 出版狀態 | Published - 2017 2月 1 |
All Science Journal Classification (ASJC) codes
- 一般醫學
指紋
深入研究「Evaluation of mechanical strengths of three types of mini-implants in artificial bones」主題。共同形成了獨特的指紋。引用此
- APA
- Author
- BIBTEX
- Harvard
- Standard
- RIS
- Vancouver