N2 - While several different methods have been used to measure carpal kinematics, biplanar radiography is generally considered to be the most accurate and popular one. However, biplanar radiography is tedious and so only pseudo-dynamic kinematics can be measured. Recently, magnetic tracking system has been developed for the measurement of joint kinematics which is versatile and easy to use and so the possibility of measuring motions dynamically. In this study, the capability of a magnetic tracking device to accurately measure carpal kinematics was investigated by comparing it with biplanar radiography. The kinematics of the third metacarpal, scaphoid and lunate in five fresh cadaveric specimens were measured using both methods as the wrists were placed in eight positions. The finite screw rotation of each bone with respect to the distal radius during selecting the seven wrist motions was calculated for both measuring techniques and compared. In general, the kinematics for all three bones measured by using either magnetic tracking device or biplanar radiography was identical and showed no statistical difference. The averaged differences ranged from 0.0 to 2.0. These differences were due to the potential effect of the weight of the sensors and the interference of the attaching rod to the surrounding tissue. It is concluded that the application of the magnetic tracking device to carpal kinematics is warranted, if proper technical procedures as suggested are followed.
AB - While several different methods have been used to measure carpal kinematics, biplanar radiography is generally considered to be the most accurate and popular one. However, biplanar radiography is tedious and so only pseudo-dynamic kinematics can be measured. Recently, magnetic tracking system has been developed for the measurement of joint kinematics which is versatile and easy to use and so the possibility of measuring motions dynamically. In this study, the capability of a magnetic tracking device to accurately measure carpal kinematics was investigated by comparing it with biplanar radiography. The kinematics of the third metacarpal, scaphoid and lunate in five fresh cadaveric specimens were measured using both methods as the wrists were placed in eight positions. The finite screw rotation of each bone with respect to the distal radius during selecting the seven wrist motions was calculated for both measuring techniques and compared. In general, the kinematics for all three bones measured by using either magnetic tracking device or biplanar radiography was identical and showed no statistical difference. The averaged differences ranged from 0.0 to 2.0. These differences were due to the potential effect of the weight of the sensors and the interference of the attaching rod to the surrounding tissue. It is concluded that the application of the magnetic tracking device to carpal kinematics is warranted, if proper technical procedures as suggested are followed.
UR - http://www.scopus.com/inward/record.url?scp=0031278535&partnerID=8YFLogxK
UR - http://www.scopus.com/inward/citedby.url?scp=0031278535&partnerID=8YFLogxK
U2 - 10.1016/S0021-9290(97)00097-3
DO - 10.1016/S0021-9290(97)00097-3
M3 - Article
C2 - 9456389
AN - SCOPUS:0031278535
SN - 0021-9290
VL - 30
SP - 1183
EP - 1186
JO - Journal of Biomechanics
JF - Journal of Biomechanics
IS - 11-12
ER -