Fracture of the femoral component associated with polyethylene wear and osteolysis after total knee arthroplasty

Chun Hsiung Huang, Chyun Yu Yang, Cheng Kung Cheng

研究成果: Article同行評審

54 引文 斯高帕斯(Scopus)

摘要

Fracture of the femoral component associated with polyethylene wear and osteolysis after total knee arthroplasty (TKA) has not been well reported before. A 63-year-old man with osteoarthritis of the right knee underwent TKA with a New Jersey LCS Knee, with cementing on the tibia and patella but not on the femoral component. After 42 months, in addition to wearing of polyethylene of the tibia and patella, severe osteonecrosis of the medial femoral condyle was noted. Osteonecrosis caused loss of osseous support of the medial flange of the femoral component, and the bone ingrowth of the central and lateral flange to the distal femur was so good that it overcame the yield stress of the metal of the femoral component and caused fracture of the femoral component. The osteolytic area was filled with autogenous iliac bone, and a new femoral component was inserted and cemented. The patient's condition became satisfactory with relief of pain. Although uncommon, fracture of the femoral component does occur associated with polyethylene wear and osteolysis.

原文English
頁(從 - 到)375-379
頁數5
期刊Journal of Arthroplasty
14
發行號3
DOIs
出版狀態Published - 1999 4月

All Science Journal Classification (ASJC) codes

  • 骨科和運動醫學

指紋

深入研究「Fracture of the femoral component associated with polyethylene wear and osteolysis after total knee arthroplasty」主題。共同形成了獨特的指紋。

引用此