Hemodynamic and metabolic effects of para- versus intraaortic counterpulsatile circulation supports

Pong Jeu Lu, Pao Yen Lin, Chi Fu Jeffrey Yang, Chun Hao Hung, Ming Yao Chan, Tzu Cheng Hsu

研究成果: Article同行評審

9 引文 斯高帕斯(Scopus)

摘要

Despite the success of intraaortic balloon counterpulsation, data on physiologic indices and optimal inflation/deflation timing control of chronic counterpulsation devices are unclear. This study explored the acute hemodynamic and metabolic efficacy of a novel 40-ml stroke volume paraaortic blood pump (PABP) versus a standard intraaortic balloon pump (IABP). Acute porcine model was used with eight pigs randomly divided into PABP (n = 4) and IABP (n = 4) groups. Hemodynamic and metabolic measurements were obtained with and without mechanical assistance. In one pig, the inflation/deflation control was adjusted to different settings, with corresponding performance indices measured. The PABP significantly improved classical counterpulsation indices (p ≤ 0.05) and achieved an average beneficial effect on these indices 1.5-3.5 times greater than that of the IABP. Classical metabolic indices (tension time index and endocardial viability ratio [EVR]), and indices new to chronic counterpulsation research (coronary perfusion, left ventricular stroke work (SW), and a newly derived EVR) were also used in assessment. Both IABP assistance and PABP assistance improved these physiologic indices, with a trend toward PABP superiority in reducing left ventricular SW (p = 0.08). An optimal PABP deflation timing occurs during systole (25 milliseconds after the R-wave) and can minimize coronary regurgitation.

原文English
頁(從 - 到)19-25
頁數7
期刊ASAIO Journal
57
發行號1
DOIs
出版狀態Published - 2011 一月 1

All Science Journal Classification (ASJC) codes

  • 生物物理學
  • 生物工程
  • 生物材料
  • 生物醫學工程

指紋

深入研究「Hemodynamic and metabolic effects of para- versus intraaortic counterpulsatile circulation supports」主題。共同形成了獨特的指紋。

引用此