Investigation of mouse conductance catheter position deviation effects on volume measurements by finite element models

Chia Ling Wei, Po Yi Wu

研究成果: Conference contribution

3 引文 斯高帕斯(Scopus)

摘要

The conductance catheter system is used to measure the instantaneous ventricular conductance, and real-time ventricular volumes is then determined by converting the measured conductance to volume. In fact, two different conductance-to-volume conversion equations for conductance catheters have been proposed, the Baan's classic equation and Wei's nonlinear equation. The accuracy of this volume estimation method is limited by several factors, such as the deviation of the catheter position inside the ventricle. The effects of the mouse catheter radial and longitudinal position deviations on the measured conductance are investigated with finite element models. Moreover, the capacities of the two conversion equations to calibrate the error induced by the catheter position variation are evaluated and compared. According to the simulation results, the error-calibrated capacity of the nonlinear conversion equation is better.

原文English
主出版物標題Proceedings of the 30th Annual International Conference of the IEEE Engineering in Medicine and Biology Society, EMBS'08
頁面1399-1402
頁數4
出版狀態Published - 2008 12月 1
事件30th Annual International Conference of the IEEE Engineering in Medicine and Biology Society, EMBS'08 - Vancouver, BC, Canada
持續時間: 2008 8月 202008 8月 25

出版系列

名字Proceedings of the 30th Annual International Conference of the IEEE Engineering in Medicine and Biology Society, EMBS'08 - "Personalized Healthcare through Technology"

Other

Other30th Annual International Conference of the IEEE Engineering in Medicine and Biology Society, EMBS'08
國家/地區Canada
城市Vancouver, BC
期間08-08-2008-08-25

All Science Journal Classification (ASJC) codes

  • 電腦視覺和模式識別
  • 訊號處理
  • 生物醫學工程
  • 健康資訊學

指紋

深入研究「Investigation of mouse conductance catheter position deviation effects on volume measurements by finite element models」主題。共同形成了獨特的指紋。