Kinematics of the proximal interphalangeal joint of the finger after surface replacement

S. Uchiyama, W. P. Cooney, R. L. Linscheid, G. Niebur, K. N. An

研究成果: Article同行評審

39 引文 斯高帕斯(Scopus)

摘要

Nine fresh-frozen normal human cadaveric long fingers were used to corn pare the kinematics of the proximal interphalangeal joint (PIP) before and after a resurfacing metal-polyethylene prosthetic replacement (Avanta prosthesis, San Diego, CA) using the magnetic Isotrak system (Polhemus Navigational Systems, Colchester, VT). The kinematics of the PIP joint after replacement were similar to that of the normal joint. The maximum angular displacement was 5°for lateral deviation and 9°for rotation during the passive flexion and extension motion. The center of rotation after implant insertion was nearly identical to the center of rotation of the normal joint. This anatomically designed Pip prosthesis has potential to restore normal motion to the finger PIP joint while resisting physiologic out-of-plane forces such as pinch and grasp. (C) 2000 by the American Society for Surgery of the Hand.

原文English
頁(從 - 到)305-312
頁數8
期刊Journal of Hand Surgery
25
發行號2
DOIs
出版狀態Published - 2000 3月

All Science Journal Classification (ASJC) codes

  • 手術
  • 骨科和運動醫學

指紋

深入研究「Kinematics of the proximal interphalangeal joint of the finger after surface replacement」主題。共同形成了獨特的指紋。