摘要
An analog front end for diagnosing and monitoring the behavior of the heart is presented. This sensing front end has two low-power processing devices, including a 5(th)-order Butterworth operational transconductance-C (OTA-C) filter and an 8-bit successive approximation analog-to-digital converter (SAADC). The components fabricated in a 0.18-microm CMOS technology feature with power consumptions of 453 nW (filter) and 940 nW (ADC) at a supply voltage of 1 V, respectively. The system specifications in terms of output noise and linearity associated with the two integrated circuits are described in this paper.
| 原文 | English |
|---|---|
| 頁(從 - 到) | 1683-1686 |
| 頁數 | 4 |
| 期刊 | Conference proceedings : ... Annual International Conference of the IEEE Engineering in Medicine and Biology Society. IEEE Engineering in Medicine and Biology Society. Conference |
| 出版狀態 | Published - 2008 |
All Science Journal Classification (ASJC) codes
- 訊號處理
- 生物醫學工程
- 電腦視覺和模式識別
- 健康資訊學
指紋
深入研究「Low-power signal processing devices for portable ECG detection.」主題。共同形成了獨特的指紋。引用此
- APA
- Author
- BIBTEX
- Harvard
- Standard
- RIS
- Vancouver