Measurement of local pulse wave velocity for carotid artery by using an ultrasound-based method

Chieh Ju Tang, Po Yang Lee, Yi Hsiang Chuang, Chih Chung Huang

研究成果: Article同行評審

10 引文 斯高帕斯(Scopus)

摘要

Currently, pulse wave velocity (PWV) is an important physical index for characterizing the mechanical properties of arteries. Carotid-femoral PWV (cfPWV) is a clinically-approved parameter for evaluating the cardiovascular risk and therapeutic efficacy. However, cfPWV only provides global information about vessel properties. Many recent studies have indicated that local PWV measurements provide precise evaluation of artery conditions. Here, an ultrasound (US) method based on a novel vessel displacement waveform correction, is proposed for improving the accuracy of local carotid PWV measurement. A programmable US device and a commercial array transducer were used, which allow a user to excite transducer and receive US signals arbitrarily with different beam settings. The local PWV measurement accuracy was verified using a phantom. The number of US beams used for PWV measurements was also considered, which indicates that eight elements is the acceptable setting. Subsequently, local carotid PWV and cfPWV were measured in 35 healthy human subjects (age: 21.9 ± 2.4 years) by using the US method and SphygmoCor device, respectively. The cfPWV and local carotid PWV were 6.65 ± 0.74 and 4.63 ± 0.57 m/s, respectively. A good linear correlation was observed between the two aforementioned methods (r = 0.8) for the subjects. All the results indicated that when few US beams were used, the proposed method exhibited a reliable measurement of local PWV.

原文English
文章編號106064
期刊Ultrasonics
102
DOIs
出版狀態Published - 2020 3月

All Science Journal Classification (ASJC) codes

  • 聲學與超音波

指紋

深入研究「Measurement of local pulse wave velocity for carotid artery by using an ultrasound-based method」主題。共同形成了獨特的指紋。

引用此