Quantitative assessment of first annular pulley and adjacent tissues using high-frequency ultrasound

研究成果: Article

1 引文 (Scopus)

摘要

Due to a lack of appropriate image resolution, most ultrasound scanners are unable to sensitively discern the pulley tissues. To extensively investigate the properties of the A1 pulley system and the surrounding tissues for assessing trigger finger, a 30 MHz ultrasound system was implemented to perform in vitro experiments using the hypodermis, A1 pulley, and superficial digital flexor tendon (SDFT) dissected from cadavers. Ultrasound signals were acquired from both the transverse and sagittal planes of each tissue sample. The quantitative ultrasonic parameters, including sound speed, attenuation coefficient, integrated backscatter (IB) and Nakagami parameter (m), were subsequently estimated to characterize the tissue properties. The results demonstrated that the acquired ultrasound images have high resolution and are able to sufficiently differentiate the variations of tissue textures. Moreover, the attenuation slope of the hypodermis is larger than those of the A1 pulley and SDFT. The IB of A1 pulley is about the same as that of the hypodermis, and is very different from SDFT. The m parameter of the A1 pulley is also very different from those of hypodermis and SDFT. This study demonstrated that high-frequency ultrasound images in conjunction with ultrasonic parameters are capable of characterizing the A1 pulley system and surrounding tissues.

原文English
文章編號107
期刊Sensors (Switzerland)
17
發行號1
DOIs
出版狀態Published - 2017 一月 7

指紋

pulleys
Pulleys
hypodermis
flexors
Ultrasonics
Subcutaneous Tissue
tendons
Tissue
Tendons
image resolution
ultrasonics
Cadaver
attenuation coefficients
Fingers
Image resolution
scanners
textures
Textures
actuators
attenuation

All Science Journal Classification (ASJC) codes

  • Analytical Chemistry
  • Atomic and Molecular Physics, and Optics
  • Biochemistry
  • Instrumentation
  • Electrical and Electronic Engineering

引用此文

@article{370e0bc551a4442fb6a541a95596c230,
title = "Quantitative assessment of first annular pulley and adjacent tissues using high-frequency ultrasound",
abstract = "Due to a lack of appropriate image resolution, most ultrasound scanners are unable to sensitively discern the pulley tissues. To extensively investigate the properties of the A1 pulley system and the surrounding tissues for assessing trigger finger, a 30 MHz ultrasound system was implemented to perform in vitro experiments using the hypodermis, A1 pulley, and superficial digital flexor tendon (SDFT) dissected from cadavers. Ultrasound signals were acquired from both the transverse and sagittal planes of each tissue sample. The quantitative ultrasonic parameters, including sound speed, attenuation coefficient, integrated backscatter (IB) and Nakagami parameter (m), were subsequently estimated to characterize the tissue properties. The results demonstrated that the acquired ultrasound images have high resolution and are able to sufficiently differentiate the variations of tissue textures. Moreover, the attenuation slope of the hypodermis is larger than those of the A1 pulley and SDFT. The IB of A1 pulley is about the same as that of the hypodermis, and is very different from SDFT. The m parameter of the A1 pulley is also very different from those of hypodermis and SDFT. This study demonstrated that high-frequency ultrasound images in conjunction with ultrasonic parameters are capable of characterizing the A1 pulley system and surrounding tissues.",
author = "Lin, {Yi Hsun} and Tai-Hua Yang and Shyh-Hau Wang and Fong-chin Su",
year = "2017",
month = "1",
day = "7",
doi = "10.3390/s17010107",
language = "English",
volume = "17",
journal = "Sensors",
issn = "1424-3210",
publisher = "Multidisciplinary Digital Publishing Institute (MDPI)",
number = "1",

}

TY - JOUR

T1 - Quantitative assessment of first annular pulley and adjacent tissues using high-frequency ultrasound

AU - Lin, Yi Hsun

AU - Yang, Tai-Hua

AU - Wang, Shyh-Hau

AU - Su, Fong-chin

PY - 2017/1/7

Y1 - 2017/1/7

N2 - Due to a lack of appropriate image resolution, most ultrasound scanners are unable to sensitively discern the pulley tissues. To extensively investigate the properties of the A1 pulley system and the surrounding tissues for assessing trigger finger, a 30 MHz ultrasound system was implemented to perform in vitro experiments using the hypodermis, A1 pulley, and superficial digital flexor tendon (SDFT) dissected from cadavers. Ultrasound signals were acquired from both the transverse and sagittal planes of each tissue sample. The quantitative ultrasonic parameters, including sound speed, attenuation coefficient, integrated backscatter (IB) and Nakagami parameter (m), were subsequently estimated to characterize the tissue properties. The results demonstrated that the acquired ultrasound images have high resolution and are able to sufficiently differentiate the variations of tissue textures. Moreover, the attenuation slope of the hypodermis is larger than those of the A1 pulley and SDFT. The IB of A1 pulley is about the same as that of the hypodermis, and is very different from SDFT. The m parameter of the A1 pulley is also very different from those of hypodermis and SDFT. This study demonstrated that high-frequency ultrasound images in conjunction with ultrasonic parameters are capable of characterizing the A1 pulley system and surrounding tissues.

AB - Due to a lack of appropriate image resolution, most ultrasound scanners are unable to sensitively discern the pulley tissues. To extensively investigate the properties of the A1 pulley system and the surrounding tissues for assessing trigger finger, a 30 MHz ultrasound system was implemented to perform in vitro experiments using the hypodermis, A1 pulley, and superficial digital flexor tendon (SDFT) dissected from cadavers. Ultrasound signals were acquired from both the transverse and sagittal planes of each tissue sample. The quantitative ultrasonic parameters, including sound speed, attenuation coefficient, integrated backscatter (IB) and Nakagami parameter (m), were subsequently estimated to characterize the tissue properties. The results demonstrated that the acquired ultrasound images have high resolution and are able to sufficiently differentiate the variations of tissue textures. Moreover, the attenuation slope of the hypodermis is larger than those of the A1 pulley and SDFT. The IB of A1 pulley is about the same as that of the hypodermis, and is very different from SDFT. The m parameter of the A1 pulley is also very different from those of hypodermis and SDFT. This study demonstrated that high-frequency ultrasound images in conjunction with ultrasonic parameters are capable of characterizing the A1 pulley system and surrounding tissues.

UR - http://www.scopus.com/inward/record.url?scp=85012070585&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=85012070585&partnerID=8YFLogxK

U2 - 10.3390/s17010107

DO - 10.3390/s17010107

M3 - Article

C2 - 28067854

AN - SCOPUS:85012070585

VL - 17

JO - Sensors

JF - Sensors

SN - 1424-3210

IS - 1

M1 - 107

ER -