The Biomechanical Properties of a High-Tensile Strength Tape for Tendon Graft Fixation Using the Krackow Configuration

研究成果: Article同行評審

4 引文 斯高帕斯(Scopus)

摘要

Purpose: To compare the biomechanical properties of a high-tensile strength suture with the high-tensile strength tape for tendon graft fixation. Methods: A total of 24 porcine tendons were used and were randomly divided into 2 groups. Two kinds of suture materials, a braided nonabsorbable high-strength suture (group S) and a high-tensile strength tape (group T), were used to complete 3 pairs of Krackow stitches on the tendons. Each specimen was pretensioned to 100 N for 3 cycles, cyclically loaded from 50 to 200 N for 200 cycles, and finally loaded to failure. Elongation after cyclic loading, ultimate failure load, and the mode of failure were recorded. Results: The elongation after cyclic loading between group S (26% ± 5%) and group T (24% ± 5%) were not significantly different (P =.378). The ultimate failure loads in group T (400 ± 38 N) were significantly greater than those in group S (358 ± 21 N) (P =.010). All specimens failed because of suture material breakage. Conclusions: Compared with the braided nonabsorbable high-strength suture, the high-tensile strength tape had similar elongation values after cyclic loading, but significantly greater ultimate failure load in this porcine in vitro biomechanical model. Clinical Relevance: A secure suture-tendon construct is especially important when a post-tie fixation technique is used because the mitigating construct may potentially lead to graft loosening and affect graft healing.

原文English
頁(從 - 到)e347-e351
期刊Arthroscopy, Sports Medicine, and Rehabilitation
2
發行號4
DOIs
出版狀態Published - 2020 8月

All Science Journal Classification (ASJC) codes

  • 公共衛生、環境和職業健康
  • 復健
  • 物理治療、運動療法和康復
  • 骨科和運動醫學

指紋

深入研究「The Biomechanical Properties of a High-Tensile Strength Tape for Tendon Graft Fixation Using the Krackow Configuration」主題。共同形成了獨特的指紋。

引用此