The relation between micromotion and screw fixation in acetabular cup

Jui Ting Hsu, Kuo An Lai, Qingshan Chen, Mark E. Zobitz, Heng Li Huang, Kai Nan An, Chih Han Chang

研究成果: Article同行評審

45 引文 斯高帕斯(Scopus)

摘要

One of the major causes inducing loosening in the cementless acetabular cup implanting is its insufficient initial stability. In this study, three-dimensional finite element models of the pelvis and acetabular components were developed to investigate the relationship between relative micromotion, initial stability, and screw fixation under six daily activity loadings. A commercial available hemispheric cup with five screw holes was used as the target acetabular cup. The simulation results showed that if screws were placed closed together, when the screw number increased from 1 to 5, the peak micromotion decreased less than 14%, from 126.5 to 108.8 μm, while the stable region, micromotion less than 28 μm, enlarged only by 40%, from 46.1% to 64.7%. However, if the screw could be placed near the cup rim, a single rim screw, 202.1 μm micromotion, could provide better stability than that of four dome screws, 209.6 μm micromotion, placed closed together. To conclude, multiple cup screws should be placed near cup rim and as separate as possible to enlarge the stable region and reduce the peak micromotion between cup and acetabulum.

原文English
頁(從 - 到)34-41
頁數8
期刊Computer Methods and Programs in Biomedicine
84
發行號1
DOIs
出版狀態Published - 2006 10月

All Science Journal Classification (ASJC) codes

  • 軟體
  • 電腦科學應用
  • 健康資訊學

指紋

深入研究「The relation between micromotion and screw fixation in acetabular cup」主題。共同形成了獨特的指紋。